Why is an ionic strength adjustor necessary for the ISE? Does it need to be used for calibration?

Document ID

Document ID TE9083

Published Date

Published Date 23/09/2019
Question
Why is an ionic strength adjustor necessary for the ISE? Does it need to be used for calibration?
Summary
Ion-selective IntelliCal electrode – ionic strength adjustor for measurement and calibration
Answer
Addition of the appropriate ionic strength adjustor is essential before carrying out a measurement.
This raises all the interfering parameters that are relevant to the measurement process to a predefined level, ensuring that the measuring environment is always constant.
The ionic strength adjustor is therefore used for both measurement and calibration.

 Powder pillows are available for the following parameters:
 Na:      Product no. 4451569
 Cl-:      Product no. 2318069
 F-:       Product no. 258999 or: LZW9900.99
 NH4:    Product no. 2980699
 NH3:    Product no. 4447169
 NO3:    Product no. 2984799