Is there a procedure for setting the Pocket Pro and Pocket Pro+ meters back to default factory settings?

Document ID

Document ID TE9101

Published Date

Published Date 11/04/2019
Question
Is there a procedure for setting the Pocket Pro and Pocket Pro+ meters back to default factory settings?
Summary
Default factory settings on a Pocket Pro and Pocket Pro+ meter.
Answer
The Pocket Pro and Pocket Pro+ meters can be set back to default factory settings.

The procedure is as follows:
Push and hold the "wrench" key until "SEt" shows on the display
Push the "wrench" key to scroll through the settings until "rSEt" is displayed.
Push the up or down arrow keys to change the value from "No" (default value) to "Yes". When set to "Yes", the meter will be reset to the default factory settings.
Push and hold the "wrench" key until "End" shows to go to continuous measurement mode.

A procedure is also given in the full user manual for each respective Pocket Pro and Pocket Pro+ meter.