What can I use to clean Orbisphere Luminescent Dissolved Oxygen (LDO) series sensors such as the K1x00, M1100 & the 3100?

Document ID

Document ID TE9225

Published Date

Published Date 01/10/2018
Question
What can I use to clean Orbisphere Luminescent Dissolved Oxygen (LDO) series sensors such as the K1x00, M1100 & the 3100?
Summary
Cleaning the Orbisphere LDO series sensors.
Answer
Hach recommends using Ethanol (EtOH) to clean our LDO sensors. We have validated the cleaning at the factory with 96% EtOH, so you can use up to 100% EtOH without issue. Do not forget to dry the path with air or N2 after cleaning with EtOH.

Denatured alcohol which is a mixture of ethanol and methanol is an acceptable option also.

Use of rubbing alcohol or anything that contains Isopropyl alcohol is not recommended due to possible bleaching of the spot.