How can I convert between NH4 (ammonium) to NH3 (ammonia)?

Document ID

Document ID TE8669

Published Date

Published Date 06/06/2018
Question
How can I convert between NH4 (ammonium) to NH3 (ammonia)?
Summary
NH4-N to NH4 and NH4 to NH3 conversions.
Answer
Ammonium has a chemical formula consisting of (1) Nitrogen and (4) Hydrogen molecules:
NH4
 
The molecular weight of NH4  is calculated as below:
N = 14.0067
H = 1.0079
So, the molecular weight of NH4  would be 14.0067 + (1.0079 X 4) = 18.0383
 
  • To convert the measured value of NH4 –N to NH4, use the following conversion formula:
NH4– N X 1.2878 (derived by dividing the molecular weight of NH4  by N) = NH4.
 
  • Using the equation below, you can get a pretty close approximation for converting NH4 to NH3:
NH4 X 0.9441 (derived by dividing the molecular weight of NH3  by NH4) = approximate NHconcentration.
 

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.