Microbiology Guide: Most Probable Number (MPN) Media

Document ID

Document ID TE8846

Published Date

Published Date 06/06/2018
Question
Microbiology Guide: Most Probable Number (MPN) Media
Summary
Microbiology Guide: Media and Equipment Selection; Most Probable Number (MPN) Media
Answer
From: Microbiology Guide: Media and Equipment Selection

Most Probable Number (MPN) Media Selection Summary

Drinking Water and Beverage
Organism USEPA- Accepted (Y/N) Presumptive Media Confirmation Media
Total Coliforms Y Lauryl Tryptose (LT) Broth Tubes (Concentrated) Brilliant Green Bile (BGB) Tubes
Total Coliforms and E. Coli N LT with MUG Broth Tubes (Concetrated) No Confirmation Necessary for E.Coli BGB Broth Tubes for Total Coliform Confirmation
Fecal Coliforms Y LT Broth Tubes (Concetrated) EC Medium Tubes
E. Coli Y LT Broth Tubes (Concetrated) EC Medium with MUG Tubes

Sludge, Recreational Water and Nonpotable Water
Organism USEPA- Accepted (Y/N) Presumptive Media Confirmation Media
Total Coliforms and E. Coli N LT with MUG Broth Tubes (Concentrated) No Confirmation Necessary for E.Coli BGB Broth Tubes for Total Coliform Confirmation
Fecal Coliforms Y LT Broth Tubes (Concetrated) EC Medium Tubes
Fecal Coliforms Y A-1 Medium Broth Tubes No Confirmation Necessary
Fecal Coliforms N PathoScreen Medium MPN Pillows Use Traditional Confirmation Media
Fecal Streptococci Y Azide Dextrose Broth Tubes Bile Esculin Azide Agar Tubes

Foods
Organism USEPA- Accepted (Y/N) Presumptive Media Confirmation Media
Total Coliforms N LT Broth Tubes (Single-Strength) BGB Broth Tubes
Fecal Coliforms N LT Broth Tubes (Single-Strength) EC Medium Tubes
E. Coli N LT Broth Tubes (Single-Strength) EC Medium with MUG Tubes
E. Coli N LT with MUG Broth Tubes (Single-Strength) No Confirmation Necessary for E.Coli BGB Broth Tubes for Total Coliform Confirmation

Media with Confirmation
  • Lauryl Tryptose with Brilliant Green, EC/Mug Media, Hach method 8001, Total, Fecal, and E.Coli Coliform
    • For Potable Water Samples
      • Most Probable Number Method 8001 for potable water is USEPA-accepted. Method 8001 meets or exceeds the specification criteria stated in Standard Methods for the Examination of Water and Wastewater, 19th edition, 9221 Multiple-Tube Fermentation Technique for Members of the Coliform Group. For potable water, confirm fecal coliforms with EC Medium Broth as cited in 40 CFR Part 141.21, Subpart (F)(5); or confirm E. coli with EC/MUG Medium Broth as cited in 40 CFR Part 141.21, Subpart (F)(6)(i).
    • Media Packaging
      • Total Coliform: Lauryl tryptose (LT) broth tubes (Product # 2101415)
      • Total Colifom: Brilliant green bile (BGB) broth tubes (Product # 32215)
      • Fecal Coliform: EC medium broth tubes (Product # 1410415)
      • Ecoli Coliform: EC medium with MUG  broth tubes (Product # 2282415)
        • Needed Accessory
          • UV long wave lamp, portable (Product # 2415200)
          • laboratory (115V) (Product # 2184300)
          • laboratory (240V) (Product # 2184302)
 
  • Lauryl Tryptose with Brilliant Green, EC/Mug Media, Hach method 8001A, Total, Fecal, and E.Coli Coliform
    • For Non-Potable Water and wastewater Samples
      • Most Probable Number Method 8001A for wastewater is USEPA-accepted. Method 8001A meets or exceeds the specification criteria stated in Standard Methods for the Examination of Water and Wastewater, 19th edition, 9221 Multiple-Tube Fermentation Technique for Members of the Coliform Group
    • Media Packaging
      • Total Coliform: Lauryl tryptose (LT) broth tubes (Product # 2101415)
      • Total Colifom: Brilliant green bile (BGB) broth tubes (Product # 32215)
      • Fecal Coliform: EC medium broth tubes (Product # 1410415)
      • Ecoli Coliform: EC medium with MUG  broth tubes (Product # 2282415)
        • Needed Accessory
          • UV long wave lamp, portable (Product # 2415200)
          • laboratory (115V) (Product # 2184300)
          • laboratory (240V) (Product # 2184302)
 
  • Lauryl Tryptose with MUG, Hach method 8091, Total, and E.Coli Coliform
    • For Potable and Non-Potable Water *Not USEPA accepted
    • Media Packaging
      • Lauryl tryptose w/MUG broth tubes (Product # 2217515)
      • Lauryl tryptose w/MUG, dehydrated (Product # 225573)
    • Needed Accessory
      • UV long wave lamp, portable (Product # 2415200)
      • laboratory (115V) (Product # 2184300)
      • laboratory (240V) (Product # 2184302)
 
  • A-1 medium, Hach method 8368, Fecal Coliform
    • For Non-Potable and Wastewater
      • Most Probable Number Method 8368, A-1 Medium, is USEPA-accepted for testing non-potable waters. Method 8368 meets or exceeds the specification criteria stated in Standard Methods for the Examination of Water and Wastewater, 18th edition, 9221 E. Fecal Coliform Procedure. USEPA Manual for the Certification of Laboratories Analyzing Drinking Water states “5.5.3. A-1 medium may be used as an alternative to EC Medium to enumerate fecal coliforms in source water, in accordance with the Surface Water Treatment Rule. A-1 Medium must not be used for drinking water samples.”
    • Media Packaging
      • A-1 medium broth tubes (Product # 2560915)
Media Fill Volumes
For concentrated media there is 2 mL of media in each tube.
For single-strength media there is 10 mL of media in each tube.

Media Decontamination
  • Recommendation:
    • Open each MPN tube and pipet in 2-3 mls of Clorox. (you can use a disposable plastic droppers).
    • Cap and invert …. Let stand for minimum 20-30 minutes.
    • After waiting time, use a wire strainer or colander, empty the contents of the MPN tubes in to the strainer with the TAP water running.  Rinse the tubes with TAP water.
    • The empty tubes can now be disposed of in the glass waste container. Caps can be disposed of in the regular waste can.
    • After all the tubes have been dosed, add additional amount of Clorox down the drain, ~20 mls. Run more Tap water into the sink.