What is the conversion factor between benzotriazole and tolytriazole?

Document ID

Document ID TE969

Version

Version 2.0

Status

Status Published

Published Date

Published Date 27/05/2017
Question
What is the conversion factor between benzotriazole and tolytriazole?
Summary
Conversion factor between benzotriazole (BT) and tolytriazole (TT)
Answer
The correct program should be used, rather than trying to convert between benzotriazole and tolytriazole, since the curves are nonlinear. The theoretical conversion between BT and TT is a simple weight conversion. Molecular weights are: benzotriazole 119.12 g/mol and tolytriazole 133.15 g/mol. The factor is closer to 1.2 depending on the instrument used.