Why does the preservation of a sample with acid change the results to total nitrate + nitrite?

Document ID

Document ID TE943

Published Date

Published Date 27/05/2017
Question
Why does the preservation of a sample with acid change the results to total nitrate + nitrite?
Summary
Preservation of samples for nitrate analysis
Answer
In a sample that is preserved, any nitrite that was present has been converted to nitrate. Even after neutralizing the sample, the nitrite that was originally in the sample remains as nitrate. On preserved samples, it is not possible to differentiate nitrite from nitrate, so it is not possible to perform nitrite analysis on preserved samples.