What is the procedure for the SpecCheck Gel Secondary Standard Set, Fluoride (catalog number 2712500)?

Document ID

Document ID TE859

Published Date

Published Date 27/05/2017
Question
What is the procedure for the SpecCheck Gel Secondary Standard Set, Fluoride (catalog number 2712500)?
Summary
Procedure for SpecCheck Gel Secondary Standard Set, Fluoride (catalog number 2712500)
Answer
1. Place STD 1 into the cell holder with the alignment mark facing the keypad, cover the cell with the instrument cap.
2. Press ZERO. The display will show "0.00" or "0.0" depending on the range.
3. Place the blank cell into the cell holder. Cover the cell with the instrument cap.
4. Press READ/ENTER. Record the concentration measurement.
5. Repeat steps 1-4 with cells labeled STD 2 and STD 3.