Does Hach have a test for potassium permanganate?

Document ID

Document ID TE785

Published Date

Published Date 26/08/2020
Question
Does Hach have a test for potassium permanganate?
Summary
Testing for potassium permanganate
Answer
Hach does not have a direct method for potassium permanganate. However Hach method 8034, the Periodate Oxidation Method for Manganese, can be adapted to measure permanganate. To adapt this method for permanganate, do not add the sodium periodate reagent. Add the buffer reagent as usual. The reaction time is not necessary. It is important to zero on deionized water and not the sample for this method modification. It is also recommended to use the same sample cell for both the blank and sample measurement.

Multiply the manganese results by 2.16 to get results as permanganate (MnO 4), by 2.88 for results as potassium permanganate (KMnO 4), or select the MnO 4 or KMnO 4 form display on the instrument.

This is not a highly sensitive test and may not measure below 1.5 - 2 mg/L KMnO 4 accurately.

 