Should the DR1900 Rescan 100%T function in the system checks menu be set to "on"?

Document ID

Document ID TE730

Version

Version 2.0

Status

Status Published

Published Date

Published Date 27/05/2017
Question
Should the DR1900 Rescan 100%T function in the system checks menu be set to "on"?
Summary
DR1900 Rescan 100%T function
Answer
When the Rescan 100%T is set to "on" the DR1900 will be forced to zero when a cuvette (TNT+ or LCK) program is selected. Generally the TNT+ or LCK vials do not need to take a zero, because the instrument will automatically use the air blank that was recorded when the DR1900 was manufactured. However, the optics can become dirty over time causing a shift on the zero value and inaccuracy in the measurement. The default is "off".