Do you have any experience with use of, LCK304, LCK340, LCK341 and LCK362 for analysis in seawater / brackish water?

Document ID

Document ID TE7137

Published Date

Published Date 16/07/2020
Question
Do you have any experience with use of, LCK304, LCK340, LCK341 and LCK362 for analysis in seawater / brackish water?
Summary
Usage of LCK in general for analysis in seawater / brackish water.
Answer
In general all our LCK tests are developed and tested for water and waste water samples.
Any sample application different from that has not been tested and so we cannot suggest anything on brackish water samples.
If the interference list does not state any interference with high Cl loads or other known parameter in concentrations you can find in brackish water I would ask the customer to test the suitability of the LCK for his samples with standard addition by his own.
 