What is the difference between contact and non-contact liquid detect on an AS950 sampler?

Document ID

Document ID TE7087

Version

Version 2.0

Status

Status Published

Published Date

Published Date 05/06/2017
Question
What is the difference between contact and non-contact liquid detect on an AS950 sampler?
Summary
Difference between contact and non-contact liquid detect
Answer
A standard contact liquid detect means that the sample passes the sensor directly. The advantage is an exact liquid detection, but the sensor can be damaged e.g. by gravel or stones.
A non-contact liquid detector option is available for customers
that demand that everything that comes into contact with the sample be replaced to reduce the risk of cross contamination. T the liquid pass the sensor indirectly through the suction tubing.