Question
What is the power consumption of the SC200 controller?
Summary
SC200 controller power requirements
Answer
The SC200 controller power consumption depends on whether the unit is AC or DC as well as the sensors connected to the SC200.

Sensors with higher power consumption will impact the total consumption of the device. Here are the consumption guidelines from the user manual:

AC powered controller: 100-240 VAC ±10%, 50/60 Hz; Power 50 VA with 7 W sensor/network module load, 100 VA with 28 W sensor/network module load.
24 VDC powered controller: 24 VDC—15%, + 20%; Power 15 W with 7 W sensor/network module load, 40 W with 28 W sensor/network module load.