How are the timing parameters from a brackish Lachat method with only QC8000 support data entered for the more current Omnion versions?

Document ID

Document ID TE7052

Published Date

Published Date 27/05/2017
Question
How are the timing parameters from a brackish Lachat method with only QC8000 support data entered for the more current Omnion versions?
Summary
Timing parameters from QC8000 to new software versions
Answer
The timing parameters for the QC8000 can be used for more current versions of software. They can be calculated as detailed below:

Inject to Peak Start= Inject to BW Baseline Start
Expected base width= Inject to BW Baseline End - Inject to BW Baseline Start
Brackish Water Offset= Inject to BW Integration Start - Inject to BW Baseline Start
Brackish Water Width= Inject to BW Integration End - Inject to BW Integration Start