Can the Incubator, Hach portable, 12 VDC be used while the battery is charging?

Document ID

Document ID TE6943

Published Date

Published Date 25/01/2018
Question
Can the Incubator, Hach portable, 12 VDC be used while the battery is charging?
Summary
Portable incubator battery use
Answer
The "Use the external battery pack" steps in the the user manual indicate that the Incubator, Hach portable, 12 VDC, 30 - 50 °C (± 0.5 °C; Catalog Number/Order Code 2569900) should be used with fully charged batteries. The power charger connector and the incubator power connector both connect to the same on the battery pack and so they cannot be used simultaneously.