The sensION+ meter displays SAT and 000 below it. What is wrong with the meter?

Document ID

Document ID TE693

Published Date

Published Date 13/03/2017
Question
The sensION+ meter displays SAT and 000 below it. What is wrong with the meter?
Summary
sensION+ meter displays SAT and 000 below it.
Answer
The serial number embedded in the instrument has been lost/reset, and needs to be reprogrammed. Follow the below steps:
Change the “000” on the display to “765”. After the “5” has been entered, press the check mark key to accept the code. Use the up/down arrow keys to change each digit Use the check mark key to go to the next digit Use the calibration key to go back to previous digit Once the 765 code has been entered the screen will display “SAT” and “PH”. Use the up/down arrow keys to scroll to the screen that says “SAT” and “Sn”.Press the check mark key to select the “Sn” option. You will now see a default 6 digit number in the lower section of the display.Find the 6 digit s/n on the tag on the back of the meter, and use the keypad on the meter to enter this serial number (use same keys as step 1 above).After the last digit has been entered, press the check mark key. If you entered the s/n correctly, the display will say “SAT”//”OK”, and then go back to “SAT”//”Sn”.Turn the meter off and back on.