When attempting to do a three-point calibration on a SensION+ MM150, the meter will not allow more than two points, even though the "check" button is not pressed during the calibration. What is causing this?

Document ID

Document ID TE689

Published Date

Published Date 27/05/2017
Question
When attempting to do a three-point calibration on a SensION+ MM150, the meter will not allow more than two points, even though the "check" button is not pressed during the calibration. What is causing this?
Summary
sensION+ meter does not allow for a three point calibration
Answer
The meter most likely has the newest version of firmware. The default on the new firmware is a one-point calibration, as this is best practice. The customer will need to change the number of standards in the meter settings if they need to usemore than one standard.

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.