Can the flow rate for the Orbisphere Luminescent Dissolved Oxygen (LDO) sensors be set lower than the recommended 50 mL/minute when using a flow chamber?

Document ID

Document ID TE6779

Version

Version 2.0

Status

Status Published

Published Date

Published Date 27/05/2017
Question
Can the flow rate for the Orbisphere Luminescent Dissolved Oxygen (LDO) sensors be set lower than the recommended 50 mL/minute when using a flow chamber?
Summary
Orbisphere LDO sensor minimum flow rate requirement for the flow chamber.
Answer
The LDO technology does not require sample flow in order to make a valid dissolved oxygen measurement.  The minimum flow rate of 50 mL/minute is recommended to minimize potential sample line air leak effects on low level dissolved oxygen measurements.  The flow rate should be adjusted to minimize the lag time between the sampling point, and where the sample is measured inside the flow chamber.