What is the factor to convert from PO4-P to PO4?

Document ID

Document ID TE6485

Version

Version 2.0

Status

Status Published

Published Date

Published Date 27/05/2017
Question
What is the factor to convert from PO4-P to PO4?
Summary
Converting from PO4-P to PO4
Answer
Many Hach spectophotometers and colorimeters give the option to change the chemical form of the reported value. Since this number is a ratio of the atomic masses of the elements in the chemical form, this calculation does not change the concentration of phosphorus in the sample; it only changes the value in relation to the units it is expressed in.

For example:
Changing a height from 6 feet to 72 inches, or to 183 cm, doesn’t change the actual height, it only changes the units in which it is reported. They are all mathematically equivalent.

To convert from units of orthophosphate (PO 4 3-) to phosphorus (PO 4-P), multiply the value by 0.3261. To change from phosphorus (PO 4-P) to orthophosphate (PO 43-) divide the value by 0.3261.