How is a Lachat Cadmium Column for Nitrate determination replaced?

Document ID

Document ID TE6373

Published Date

Published Date 27/05/2017
Question
How is a Lachat Cadmium Column for Nitrate determination replaced?
Summary
Cadmium column replacement
 
Answer
 
To replace a Cadmium Column for Nitrate determination (Catalog Number/Order Code 50237A) :
 
1. Remove the used cadmium column from the manifold.
2. Remove the flangeless fittings. Properly dispose of the used cadmium column
3. Remove the end plugs from the new column and install the flangeless fittings into the
column. Tighten the flangeless fittings securely.
4. Install the cadmium column on the manifold.
 