Is it possible to save a user progrom on a DR3900 or DR6000 for single wavelength absorbance measurements?

Document ID

Document ID TE6350

Published Date

Published Date 27/05/2017
Question
Is it possible to save a user progrom on a DR3900 or DR6000 for single wavelength absorbance measurements?
Summary
User program on DR3900 or DR6000 for single wavelength absorbance measurements
Answer
It is possible to save a user progrom on a DR3900 (Catalog Number/Order Code LPV440.99.00012) or DR6000 (Catalog Number/Order Code LPV441.99.00012) for single wavelength absorbance measurements.
To do this:
1) From the Main Menu, select Single Wavelength
2) Select Options
3) Select the lambda symbol (λ)
4) Key in desired wavelength
5) Select OK
6) Select Options again
7) Select More...
8) Make any desired changes to Concentration Factor, Concentration Resolution, and Reading Mode (available only on the DR6000).
9) Select Save as User Program
10) Key in desired program number
11) Select OK
12) Key in Program Name
13) Select OK

To find the newly added User Program:
1) From the Main Menu, select User Programs
2) Select program from list
3) Select Start to begin program