How long should the sample be centrifuged for Volatile acids method 8196?

Document ID

Document ID TE6259

Version

Version 2.0

Status

Status Published

Published Date

Published Date 27/05/2017
Question
How long should the sample be centrifuged for Volatile acids method 8196?
Summary
Volatile acids.
Answer
The centrifuge time required will depend upon the revolutions per minute, the length of the centrifuge arm, the g-forces applied and the consistancy of the sample. A standard time is difficult to estimate, however, the purpose of the centrifuge step is to separate the solids in sludge from the liquid portion. Based upon this, the samples need to centrifuge long enough to provide a sample (supernatant solution) which is free from solids. That should be no longer than 5 minutes. The time can be reduced or increased from there.