What do each of the valve state colors shown in Omnion correlate to on the valve?

Document ID

Document ID TE6135

Version

Version 1.0

Status

Status Published

Published Date

Published Date 13/03/2017
Question
What do each of the valve state colors shown in Omnion correlate to on the valve?
Summary
Checking valve state within Omnion
Answer
In the Timing tab of the run properties window in Omnion, small colored icons indicate what state the valve for each channel is in.  When the valve icon is red, sample should be going through the sample loop and the valve is in the load state. When the valve icon is green the valve is in the inject state. By clicking within the box you can manually change the state of the valve.  This is very helpful for determining if the valve is turning properly as well as for backflushing and cleaning the valve.