Does process concentration affect CL10 sc electrolyte usage?

Document ID

Document ID TE6113

Published Date

Published Date 27/05/2017
Question
Does process concentration affect CL10 sc electrolyte usage?
Summary
CL10 sc electrolyte usage and process concentration.
Answer
In a process with higher chlorine concentration, the readings on the CL10 sc do not remain stable for the general three to six months that is recommended between electrolyte refills.  Higher chlorine, pH, and temperature will run the electrolyte down faster so it will need to be refilled at more frequent intervals than would occur at lower chlorine concentrations.