What will happen if the warnings on the CA610 Fluoride analyser are cleared?

Document ID

Document ID TE6074

Published Date

Published Date 21/01/2019
Question
What will happen if the warnings on the CA610 Fluoride analyser are cleared?
Summary
CA610 analyser warnings.
Answer
If the user clears calibration warnings (Marg Slope, Marg Offset, or Err Slope), the analyser will give no further indication that it is running a marginal (or completely bad) calibration. 
 
If the user cycles power off/on (or power outage occurs) after occurrence of a calibration warning, the analyser will restart with the warning cleared – and it will give no further indication that it is running a marginal (or completely bad) calibration.