Can I do a custom calibration using the IntelliCAL™ CDC401 probe on the SL1000 Portable Parallel Analyzer™ (PPA)?

Document ID

Document ID TE6035

Published Date

Published Date 13/03/2017
Question
Can I do a custom calibration using the IntelliCAL™ CDC401 probe on the SL1000 Portable Parallel Analyzer™ (PPA)?
Summary
Custom calibration for conductivity on SL1000 Portable Parallel Analyzer™ (PPA)
Answer
No, custom calibrations for Conductivity are currently not available on the SL1000. Only pre-programmed standard curves can be used. See below list for available calibration points:
 
  • 1D KCl, 111.3 mS/cm, 25°C
  • 0.1D KCl, 12.85 mS/cm, 25°C
  • 0.01D KCl, 1408 µS/cm, 25°C
  • 0.1M KCl, 12.88 mS/cm, 25°C
  • 0.01M KCl, 1413 µS/cm, 25°C
  • 0.001M KCl, 146.93 µS/cm, 25°C
  • NaCl, 18 mS/cm, 25°C
  • NaCl, 1000 µS/cm, 25°C
  • NaCl, 180 µS/cm, 25°C
  • NaCl, 25 µS/cm, 25°C
  • NaCl, .05%, 1015 µS/cm, 25°C