Can a TitraLab® AT1000 with an application pack including conductivity also display results for TDS, salinity, or resistivity?

Document ID

Document ID TE5964

Published Date

Published Date 27/05/2017
Question
Can a TitraLab® AT1000 with an application pack including conductivity also display results for TDS, salinity, or resistivity?
Summary
Conductivity TitraLab® AT1000 parameters
Answer
Although these are parameters that are traditionally calculated from a measured conductivity, the TitraLab® AT1000 can only display the conductivity result from a conductivity measurement. If any of these other parameters are needed, the CDC40101 that works with the TitraLab® AT1000 will also connect to an HQ14d Portable Conductivity Meter (Catalog number/Order code HQ14D53000000) which does have the capability to calculate and display the result for these other related parameters.