After powering the instrument on, the BODTrak™II shows "Test Mode". What does this mean?

Document ID

Document ID TE5932

Published Date

Published Date 13/03/2017
Question
After powering the instrument on, the BODTrak™II shows "Test Mode". What does this mean?
Summary
Test Mode on a BODTrak™II
Answer


Test mode is used to verify operation at the factory.  To return to the normal mode, one can follow four simple steps:

1) In the measuring window, be sure the current channel is in CLEAR or END status.
2) Press 'ON' to enter SETUP status.
3) Press '1' and ''6' at the same time until "User Mode" is displayed.
4) Press 'OFF' to enter the measuring window.
 

Was this answer helpful?