What is the replacement for the obsolete Catalog Number/ Order Code 08350=A=0002 Polymetron pH sensor?

Document ID

Document ID TE5902

Published Date

Published Date 27/05/2017
Question
What is the replacement for the obsolete Catalog Number/ Order Code 08350=A=0002 Polymetron pH sensor?
Summary
Replacement for Catalog Number/ Order Code 08350=A=0002 Polymetron pH sensor.
Answer
There are two possible replacements for the Polymetron Catalog Number/ Order Code 08350=A=0002:

One is the 8350.4 (Catalog Number/ Order Code 08350=C=0004) and the other is 8350.5 (Catalog Number/ Order Code 08350=C=0005).

The correct replacement depends on the sample characteristics. If there is high temperatures and corrosive media, 8350.4 is more suitable.  If there is fouling sample and organics with more particles, 8350.5 is more suitable.