Can a relay be set to trigger upon a warning on the 5500 sc analyser?

Document ID

Document ID TE5792

Published Date

Published Date 21/01/2019
Question
Can a relay be set to trigger upon a warning on the 5500 sc analyser?
Summary
5500 sc analyser warning level settings.
Answer
In addition to standard relay setup for control, the relays can also be configured to trigger upon a warning.  In order to set the relay, take the following path:
Menu > System Setup > Set Outputs > Relay Setup > Select Relay (A, B, C, D) > Source (select the analyser name) > Function : Warning > Activation : Warning Level

Listed below are the available warnings to choose from. Any or all can be chosen.
  • High dark current
  • Reagent pressure
  • Unstable readings
  • Low light level
  • Reagent low
  • Standard solution low
  • Sample flow low
  • Heater failure
  • Fluid leak
  • Temperature sensor
  • Case temp high
  • Failed cal
  • Sample 1 missing
  • Sample 2 missing
  • Sample 3 missing
  • Sample 4 missing
  • Sample 5 missing
  • Sample 6 missing
  • Shutdown mode
  • Sample pressure > X
  • Sample pressure < X
  • Cal standard empty
  • Auto cal disabled
  • Analyser setup
  • Temp out of range
  • Open heater circuit
  • Reagent pressure high
  • Fan failure
  • Overfeed timer
  • Restart to accept SW
  • Network comm loss
  • Link2sc job failure