What is the difference between the DR3900 and DR6000?

Document ID

Document ID TE5743

Published Date

Published Date 06/03/2018
Question
What is the difference between the DR3900 and DR6000?
Summary
DR3900 vs. DR6000
Answer

The biggest different between the DR3900 and DR6000 is that the DR6000 has a UV lamp.

The DR6000 has an automatic moving carousel.

There are certain software applications that can only be used on the DR6000. These include the Brewery Analysis Package (Catalog Number/ Order Code LZV942), the Drinking Water Analysis Package (Catalog Number/ Order Code LZV925), and the Enzymatic Food Analysis Software (Catalog Number/ Order Code LZV943).

The DR6000 has tighter specifications than the DR3900. For example, the wavelength accuracy of the DR6000 is ± 1 nm. The wavelength accuracy of the DR3900 is ± 1.5 nm.
The data sheets for the DR3900 and DR6000 can both be found on the website.
 

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.