How do I set up the 'Mark End of Measure' feature on the 5500 sc analyzer?

Document ID

Document ID TE5593

Published Date

Published Date 27/05/2017
Question
How do I set up the 'Mark End of Measure' feature on the 5500 sc analyzer?
Summary
Mark end of measure feature on 5500 sc analyzer.
Answer
How to set up the Mark End of Measure on the 5500sc:
The ‘Mark End of Measure’ was pretty well hidden in the Series 5000 analyzer.  Fortunately, it’s now much easier to set up in the 5500sc:
1.  Press [Menu].
2.  Go to Setup System, then press [Enter].
3.  Go to Setup Outputs, then press [Enter].
4.  Go to Relay Setup, then press [Enter].
5.  Select relay A, B, C, or D, then press [Enter].
6.  Go to Select Source, then press [Enter].
7.  Select Analyzer_xxxxxx, then press [Enter].
8.  Go to Activation, then press [Enter].
9.  Select Select Events (no other choice is available), then press [Enter].
10.  Select Mark End of Measure, using the [<] key to mark the checkbox. 
11.  Deselect any other listed events, using the [<] key to unmark the checkboxes.  All events except Mark End of Measure MUST be deselected, or the external sequencer will not be properly triggered.
12.  Press [Home].