Question
How is the SL1000 Portable Parallel Analyzer™ software updated?
Summary
SL1000 Portable Parallel Analyzer™ software updated
Answer
The SL1000 Portable Parallel Analyzer™ can be updated with the following procedures.
1.) Plug instrument into AC power.
2.) Download the software update to PC.
3.) Connect USB cable to instrument and PC. The SL1000 Portable Parallel Analyzer™ should be automatically recognized by the PC.
4.) Drag/drop software zip file directly into instrument folder (Do not open zip file, just drop whole folder in).
5.) Safely remove USB connection.
6.) From the instrument Main Menu select Settings >Meter >Upgrade Software.

The unit will then automatically run the upgrade if found.
Verify the software update completed by selecting from the instrument Main Menu>Info>Meter and check the S/W Version matches the downloaded version.