Are both the Water Purification System Absorber and Universal Filtration Cartridges needed?

Document ID

Document ID TE5390

Version

Version 2.0

Status

Status Published

Published Date

Published Date 27/05/2017
Question
Are both the Water Purification System Absorber and Universal Filtration Cartridges needed?
Summary
Using one cartrige verses using both together.
Answer
The Water Purification System Adsorber Filtration Cartridge (Catalog number/Order code 2967000) can be used alone, but it is only recommended for filtered water.
The Water Purification System Universal Filtration Cartridge (Catalog number/Order code 2967100) can be used alone, but if the water line contains chlorine, turbidity, phosphates, chloramines, etc. the cartridge will not last long before needing to be replaced.

It's recomended that both be used together, and that the water line go through the Absorber filter before the universal filter.

See also:
What are the specifications for the Water Purification System Absorber Filtration Cartridge?
What are the specifications for the Water Purification System Universal Filtration Cartridge?