What are the specifications for the Water Purification System Absorber Filtration Cartridge?

Document ID

Document ID TE5375

Published Date

Published Date 27/05/2017
Question
What are the specifications for the Water Purification System Absorber Filtration Cartridge?
Summary
Specifications for the Water Purification System Absorber Filtration Cartridge.
Answer
The specifications for the Water Purification System Absorber Filtration Cartridge (Catalog number/Order code 2967000) are:
 
  • Grandular activated carbon
  • Pretreatment prior to D.I. cartridges
  • Removes: Organics, Turbidiy, Free Chlorine, Chloramines, and Phosphate Complexes
  • No exhaustion indication
  • Recommended to be changed out with every other cartridge change
  • 3 1/2" Diameter x 19 3/4" Length
  • 30 psi maximum rated pressure
  • 100 °F maximum temperature
  • Shelf life of 2 years
  • Design flow rate of 7.2 GPH