Is there a cheat sheet for the menu structure on the CL17?

Document ID

Document ID TE5162

Published Date

Published Date 21/01/2019
Question
Is there a cheat sheet for the menu structure on the CL17?
Summary
Menu Structure for the CL17 Chlorine analyzer.
Answer
Menu Structure
“New” CL17 Chlorine Analyser
 
Press the [Menu] key to access the following table.  [Up Arrow] and [Down Arrow] keys move up and down columns on the table.  Press [Enter] to move one column to the right in the table; press [Exit] to move one column to the left. 
 
MENU SUB MENU EDIT COMMENTS

ALARMS

A1- SA Y Will alarm if system alarm is generated
  A1- SW Y Will alarm  is system warning is generated
  A1- HI Y 0.0 mg/L to 5.0 mg/L limits. Alarms if Concentration > =TripPoint
  A1- LO Y 0.0 mg/L to 5.0 mg/L limits. Alarms if Concentration <= TripPoint
  A2- SA Y Will alarm if system alarm is generated
  A2- SW Y Will alarm  is system warning is generated
  A2- HI Y 0.0 mg/L to 5.0 mg/L limits. Alarms if Concentration > =TripPoint
  A2- LO Y 0.0 mg/L to 5.0 mg/L limits. Alarms if Concentration <= TripPoint
  RECALL WARNINGS   Momentarily displays all active warnings then displays DONE
  CLEAR WARNINGS   Clears all warnings and momentarily displays OKAY.

RECRDR

REC LO

Y 0.0 mg/L to 5.0 mg/L limits  (Rec Lo =Rec Hi -> 0, 5.0)
  REC HI Y 0.0 mg/L to 5.0 mg/L limits  (Rec Lo =Rec Hi -> 0, 5.0)
  ON SYSTEM ALARM
 
Y On System Alarm recorder fails [ LO, HI  HOLD)

MAINT

PRIME   Allows 39 sequential non-stop measurement cycles in order to PRIME the reagent lines.  EXIT key will abort the PRIME mode.
  CLEAN   Stops the Measurement Cycle with sample and reagents in the sample cell.  Will remain in the CLEAN mode for 60 minutes or until the EXIT key is pressed.
  HOLD OUTPUTS Y Follows Edit Rules.  If Alarm Lockout is enabled by selecting ON and pressing the ENTER key, all alarms are disabled and the ALARM LED flashes to remind the user that alarms have been disabled.  The alarm lockout period is 60 minutes after which the alarm lockout will be disabled.  Pressing the EXIT key will also disable the alarm lockout.
Edit Rules : Edit allows you to changes displayed values or automatically changes values when you press the valid  UP, DOWN, LEFT, RIGHT keys.  The valid arrow keys are displayed in the LCD display.    [ENTER accepts changes, EXIT ignores them and leaves old status intact].
 
 
 

 
 
MENU SUB MENU EDIT COMMENTS [default values in red]

SETUP

SIGAVG Y Select from 1, 2, 3 measurement average using the UP, DOWN keys.  Follows the Edit Rules.  Concentration is the average of the last N measurements where N= 1 to 3 as selected.
  Cal Zero Y Displays current value.  If value is <0.05 mg/L it can be set to zero by pressing the ENTER key.  Otherwise ERROR will be momentarily displayed.  MARGINAL ZERO warning will be generated if the ZERO value is >0.3 mg/L
  Cal Std Y Displays current value. If value is legitimate Edit rules apply otherwise the ENTER key will result in a momentary ERROR display.  If the value may be edited, EDIT VALUE will be displayed.  Change the displayed number to the ‘correct’ value and press the ENTER key.  If the resulting GAIN factor required to correct the displayed value is >0.5 and < 2, the GAIN will be updated, otherwise ERROR will be momentarily displayed.  MARGINAL GAIN  warning will be generated if the GAIN value is >1.3  or < 0.7
  cal defaults   Sets Gain = 1.0 and Zero = 0.  Concentration = (Measured Concentration -Zero) * Gain. 
  OFFSET   Momentarily displays the present ZERO correction factor.
  GAIN   Momentarily displays the present GAIN correction factor
  Display TEST   All LCD segments momentarily turn on
  RECMIN   Set recorder output to 4ma
  RecMAX   Set recorder output to 20ma
  Rec Adjust Y Use LEFT, RIGHT keys to select RECMIN or RECMAX then use UP, DOWN keys to change the value.  Follows the edit rules.
  Relay Test   Momentarily turns relays off, then turns on ALARM1 then turns on ALARM2 and then clears both alarms..
  Ref   Momentarily displays the last Reference A/D counts
  Sample   Momentarily displays the last Sample A/D counts
  Zero   Momentarily displays the last Zero A/D counts
  Cycle Time   Momentarily displays the Motor on time in seconds
  Power Fail Warning Y ON / OFF
  DIAG  OUTPUT Y ON / OFF
  Keyboard lockout Y Follows Edit Rules.  If Keyboard Lockout is enabled by selecting ON and pressing the ENTER key, no editing function is allowed.  In addition all diagnostics are disabled except this one.  Default is OFF
  Language Y English, German, French, Spanish     [Edit rules]
  default setup   Sets instrument default values.  All user settings will be lost!!
Edit Rules : Edit allows you to changes displayed values or automatically changes values when you press the valid UP, DOWN, LEFT, RIGHT keys.  The valid arrow keys are displayed in the LCD display.    [ENTER accepts changes, EXIT ignores them and leaves old status intact].
 

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.