Which type of Orbisphere Ozone sensors are compatible with which type of Orbisphere controllers.

Document ID

Document ID TE5019

Published Date

Published Date 21/01/2019
Question
Which type of Orbisphere Ozone sensors are compatible with which type of Orbisphere controllers.
Summary
Orbisphere ozone sensor compatibility.
Answer
Sensor types (most common)
PN 31311.12 - (Obsolete, old style, non cartridge) Polyether Ether Keytone (PEEK) body material, used on all versions of ozone controllers.
PN 31331.15 - (Obsolete, old style, non cartridge) Stainless Steel (SS) body material, ONLY used on PN 26506.xx, Residual Ozone System (ROS) ozone controllers which are also obsolete.
PN 31330.15 - (old style, non cartridge) SS body, used on all 36xx series ozone controllers as well as new style 410C/510Cxx series ozone controllers.
C1100-x00 - (current style, cartridge type) used on all 36xx series ozone controllers as well as new style 410C/510Cxx series ozone controllers.

Controller types (Identification)
26506.xx - (Obsolete, old style) ROS, SS box with Calibration knob and toggle switch.
36xx/3xx - (Obsolete, old style) Microprocessor One-Channel Analyser (MOCA -orange or gray box) or Wall logger (in-Line One-Channel Analyser (LOCA), 3660/3xx, SS box  analyzers. Must be configured for Ozone measurement.
410C/510Cxx series - (current style) touch screen.  Must be configured for Ozone measurement.

FAQ

Q: Will the 31331.15 sensor work on any controller other than the 26506.xx (ROS)?
A:  No - sensor has specific thermistor configuration for the ROS.

Q: Why is the PN 31311.12 listed as able to work with all controllers?
A: The 31311.12 is the ONLY sensor with dual thermistor configurations, allowing it to work with all controllers.