Will scratches on the vial left by the TU5300 sc/TU5400 sc Automatic Cleaning Module (ACM) influence turbidity readings?

Document ID

Document ID TE4082

Published Date

Published Date 27/05/2017
Question
Will scratches on the vial left by the TU5300 sc/TU5400 sc Automatic Cleaning Module (ACM) influence turbidity readings?
Summary
TU5300 sc/TU5400 (TU5) sc ACM maintenance information.
Answer
Use the following guidelines for whether or not scratches will affect measurements on a TU5 process turbidimeter:
-If the bottom of the vial is dirty/scratched but the vial side walls are clean, there is negligible to no influence on measurement
-If the bottom of the vial is clean but the vial side walls are dirty/scratched, there is negligible to no influence on measurement
-If both the vial bottom and vial walls are affected by dirt and scratches, turbidity will begin to rise
 

Was this answer helpful?

Submitting...
Thank you for your feedback!
There was an error with your submission. Please try again.