What is the power consumption of the TU5300 sc/TU5400 sc?

Document ID

Document ID TE3779

Published Date

Published Date 27/05/2017
Question
What is the power consumption of the TU5300 sc/TU5400 sc?
Summary
Power usage of TU5300/TU5400 sc turbidimeters.
Answer

Power consumption is independent of model (power consumption for TU5300 = power consumption for TU5400):

TU5300/TU5400 sc models: 8 W per unit

TU5300/TU5400 sc models with Automatic Cleaning Module: 12.5 W per unit

The TU5300 sc/TU5400 sc Turbidimeters can be connected to sc controllers in the following configurations:

sc200

Without automatic cleaning module: 2 per controller with max cable length of 50 m.

With automatic cleaning module: 2 per controller with max cable length of 1.5 m for any instrument with automatic cleaning module.

 

sc1000

Without automatic cleaning module: 5 per probe module with max cable length of 50 m.

With automatic cleaning module: 5 per probe module with max cable length of 1.5 m for any instrument with automatic cleaning module.