What are the specifications of the Lamp, Ultraviolet, Long-Wave 115 Vac, 60 Hz and 230 Vac, 50 Hz?

Document ID

Document ID TE3677

Published Date

Published Date 27/05/2017
Question
What are the specifications of the Lamp, Ultraviolet, Long-Wave 115 Vac, 60 Hz and 230 Vac, 50 Hz?
Summary
Ultraviolet, Long-Wave 115 Vac, 60 Hz and 230 Vac, 50 Hz specifications
Answer
Specifications of the Ultraviolet, Long-Wave 115 Vac, 60 Hz and 230 Vac, 50 Hz (Catalog Number/Order Code 2184300 and 2184302).

The manufacture UVP® has provided the following specifications for their lamp model number F6T5BL:

These lamps emit highly Concentrated radiation between 315 nm and 400 nm (at peak 352 nm) Other specifications are similar to those of general fluorescent lamps for illumination.
Lamp Wattage: 4 W
Demension: Length= 210.5 mm; Diameter= 15.5 mm
Cap: G5
Lamp Current: 0.147 A
Ultraviolet Output: 0.5 W
Total ammount of the ultraviolet rays between 315 nm and 400nm.
Average Useful Life: 3,000 Hrs
When the ultraviolet ray output becomes 60% of the initial output (after 100 Hrs of lighting).
No.60081 IEC: 1030

Warning: The radiation of these lamps is harmful to eyes and skin. Protect your eyes and skin against the ultraviolet rays.
Attachments