How do you change the temperature unit and buffer set on the H-Series miniLab pocket pH Meters™?

Document ID

Document ID TE2762

Version

Version 2.0

Status

Status Published

Published Date

Published Date 27/05/2017
Question
How do you change the temperature unit and buffer set on the H-Series miniLab pocket pH Meters™?
Summary
Instructions to change the temperature unit and buffer set on the H-Series miniLab pocket pH Meters™.
Answer
  • Start with the meter turned off.
  • While holding the CAL 1 button down, press and release the Power button while still holding the CAL 1 down.
  • The tester will flash through the buffers of the current set in the upper section and display the current temperature unit in the lower section.
  • You can press the CAL 1 to cycle through the 4 options.
    • pH 4-7-10/°C
    • pH 4-7-10/°F
    • pH 4-6.86-9.18/°C
    • pH 4-6.86-9.18/°F
  • Once you have the desired setting displayed on the screen, press the Power button. When the instrument is powered back on the settings will be changed.