How do you setup the sc1000 controller to send an email in the event an error occurs?

Document ID

Document ID TE2476

Published Date

Published Date 12/03/2017
Question
How do you setup the sc1000 controller to send an email in the event an error occurs?
Summary
sc1000 and email set up
Answer
SC1000 Email Setup
1. In the main menu go to SC1000 setup> EMAIL. So far yahoo is the only email service I have
found that works. Sender must match the email unsername. The yahoo SMTP server is
smtp.mail.yahoo.com. Enter the username and password of the account.



2. A user can set it up to email up to four accounts. Select E-MAIL 1 enter the email address. Repeat is
how often you want it to send the same email. Start time is how long it takes after an error or
warning to send the first email. Set to 0 if you want it to send immediately.


3. Go into the configure menu to choose what warnings or errors will prompt an email to be sent.
Click on add, choose sensor, sc1000 or relay. Then select the sensor, choose ERROR and/or
WARNINGS. It automatically selects all options. Uncheck any warnings or errors you do not want
to send an email for.