How do I select my trip point on the SP510 analyzer?

Document ID

Document ID TE2386

Published Date

Published Date 27/05/2017
Question
How do I select my trip point on the SP510 analyzer?
Summary
SP510 and trip point selection
Answer
The SP510 reagents can be ordered in different ranges. The alarm trip point selected should be 40 to 50% higher than the normal effluent hardness.
Alarm Trip Buffer Indicator
Point
0.3 mg/L 27685-49 27946-49
1 mg/L 27685-49 27690-49
2 mg/L 27685-49 27691-49
5 mg/L 27685-49 27692-49
10 mg/L 27686-49 27692-49
20 mg/L 27687-49 27692-49
50 mg/L 27688-49 27692-49
100 mg/L 27689-49 27692-49