How do I sample using the ultrasonic sensor durring a high alarm period with the AS950?

Document ID

Document ID TE2324

Published Date

Published Date 12/03/2017
Question
How do I sample using the ultrasonic sensor durring a high alarm period with the AS950?
Summary
AS950 and ultrasonic sensor sampling
Answer
For programming the sampler:
1 Start a datalogger program for the ultrasonic so it is recording the readings.
2 Add an alarm for the ultrasonic that is the reading that will trigger the sampler (high flow rate for starting sampler)
3 In the sampler program, go to "Pacing" then choose "Event". In the Event selection choose the Ultrasonic alarm that you made.
Once this is done the sampler will start when the "alarm" (ultrasonic) is triggered.