How do you reset the profile counter on the UVAS sc sensor?

Document ID

Document ID TE2319

Published Date

Published Date 12/03/2017
Question
How do you reset the profile counter on the UVAS sc sensor?
Summary
UVAS and reset counter
Answer
When you change the profile, you should go in the Sensor Setup Menu > Select the sensor >Test/Maint >Maint. Proc. >Replace profile. The wiper will then be driven out so you can replace the profile and display "Replace Profile!". Once you have replace the profile (wiper) press Enter. The wiper will go back in and will take you back to the Test/Maint menu, press the back key, controller will display "Return Probe to Process", press enter. "Please wait" and a status bar will be displayed. When it is done you will be taken back to the main measure screen and the counter will be reset.