How does a very cold sample temperature affect the accuracy of chlorine measurement of the CL10?

Document ID

Document ID TE2299

Version

Version 2.0

Status

Status Published

Published Date

Published Date 27/05/2017
Question
How does a very cold sample temperature affect the accuracy of chlorine measurement of the CL10?
Summary
CL10sc and temperature affects
Answer
The specifications for the CLF10 and CLT10 state the following requirements regarding sample water temperature:

Chlorine Sensor Sample Temperature: 5 to 45°C
Panel Operating Temperature: 0 to 50°C

The term “Panel” includes: SS Panel, Gateway, Chlorine Sensor Flow Cell, and the pH Flow Cell)

The chlorine sensor will continue to operate and provide measurements across the entire operating range, but if the temperature is below 5°C there will be a warning about temperature being out of range. If there is a warning condition present, Hach cannot guarantee accuracy; the measurement accuracy may or may not be within the published specifications in this case.