What is the leak test on the Series 5000 analyser? How do you do it?

Document ID

Document ID TE2221

Published Date

Published Date 21/01/2019
Question
What is the leak test on the Series 5000 analyser? How do you do it?
Summary
Series 5000 and leak test
Answer
1. Shut off power to the analyser (keep the sample line and the reagent bottles pressurized).
2. Remove the sample cell cover from the sample cell and dry off all straws.
3. Let the sample cell cover sit above a dry paper towel for at least 15 minutes. How much time did you watch for dripping?
4. Are any straws dripping? If yes replace the corresponding valve.