What should be done if Series 5000 showing CLEAN SAMPLE CELL and REPLACE LAMP warnings?

Document ID

Document ID TE2220

Published Date

Published Date 27/05/2017
Question
What should be done if Series 5000 showing CLEAN SAMPLE CELL and REPLACE LAMP warnings?
Summary
Series 5000 and sample cell and lamp warnings
Answer
1. IS THE LAMP WORKING?
     a) It will not be on during the entire cycle, but should turn on periodically.
     b) If the bulb has dark glass, replace it

2. Clean the sample cell
3. Check for obstruction in the light path:
     a) Check and clean the lens.
     b) Check the interference filter for scratches, pitting discoloration and check the function of the photocell assembly using a multi-meter on mV DC settings:
          i. remove the photocell from the colorimeter and disconnect the leads from terminal 24 and 25
          ii. In light the photocell should read around 300mV
          iii. In darkness the photocell should read around 0mV
4. Check for oxidized, corroded or loose connections between the control box and the lamp assembly, stir bar and photocell assembly. If you notice corrosion or oxidation, please clean this and make sure connections are secure.
5. How does the unit calibrate?