Service Reminders are displayed when counters expire on 5500sc. How can I reset after service is done?

Document ID

Document ID TE2143

Published Date

Published Date 11/04/2019
Question
Service Reminders are displayed when counters expire on 5500sc. How can I reset after service is done?
Summary
5500sc Reminders
Answer
When counters expire, reminders will display.
To reset the counters, do the following:
  • Hit DIAG key
  • Move down until SERVICE is highlighted and hit ENTER
  • Highlight SERVICE PART and hit ENTER
  • Highlight the reminder you want to reset and hit ENTER
  • It will show a screen with last and next service, days remaining and RESET SERVICE DATE?
  • At Bottom of screen it will have choice of NO and YES. Check in front of YES and hit ENTER