What do you multiply the offset by when performing a Solitax turbidity calibration?

Document ID

Document ID TE2133

Published Date

Published Date 11/04/2019
Question
What do you multiply the offset by when performing a Solitax turbidity calibration?
Summary
Solitax sc Turbidity Calibration
Answer
Negative one (-1). This value is missing from the current User Manual. The reason is that when you put the probe into the DI water your goal is to make the probe read zero. Therefor, if the probe is reading 5 NTU in DI water you will make an offset of -5. Then the probe will read zero in the DI water.